Laplace-Domain Beamforming for Ultrafast Plane-Wave Imaging

Abstract

Fourier-domain beamforming methods, such as filtered backpropagation (FBP), are beneficial in ultrafast plane-wave imaging due to their computational efficiency and high image quality. These methods use the Fourier diffraction theorem (FDT) for steered plane waves to exactly invert linear scattering models. However, these models rely on simplifying assumptions that degrade image quality. The use of linear transducer arrays, in particular, requires: (i) a two-dimensional space, and (ii) lossless tissues. Herein, Laplace-domain beamforming is proposed. This method retains the computational efficiency of the Fourier-domain methods while improving image quality for linear arrays. Specifically, the method uses a generalization of the FDT, here called Laplace diffraction theorem, to account for three spatial dimensions, finite array element heights, and lossy tissues. A phantom experiment showed that the proposed method improves image uniformity, spatial resolution, and contrast compared to FBP. Using three steering angles, the lateral and axial -6 dB-widths of wires reduced by up to 12% and 25%, respectively, while the contrast of anechoic regions improved by up to 9%.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Martin F. Schiffner. 2026-10-06. Laplace-Domain Beamforming for Ultrafast Plane-Wave Imaging. https://arxiv.org/abs/2610.08080

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

A No-Go Framework for Coarse-Grained Effect Prediction in Boron Neutron Capture Therapy: Structural Identifiability Limits and Microscopic Coverage

We establish when coarse-grained observables uniquely determine an expectation-based biological response in boron neutron capture therapy (BNCT). Microscopic target populations are represented by probability distributions, while retained observables are finitely many population-averaged linear descriptors. For an admissible class containing all finitely supported microscopic distributions, exact distribution-free identification holds if and only if the single-target response kernel lies in the span of the constant function and retained descriptors. Thus mean dose, mixed-field component means, and finite moment sets are not generally sufficient under unrestricted microscopic heterogeneity. We also derive irreducible ambiguity and approximation bounds, sharp partial-identification intervals for bounded non-increasing convex survival kernels, a treatment-ranking criterion, and lower-tail and high-exposure bounds. A fixed-total-exposure temporal model distinguishes transient re-sampling from persistent target-specific underexposure, and a marked-Poisson model shows that separate frequency and event-quality marginals need not determine survival. The framework clarifies which microscopic information is lost through BNCT coarse-graining while leaving room for approximation, partial identification, and prediction under independently justified model restrictions.

physics.med-ph↗

A Multimodal Hyperspectral and Laser Speckle Imaging Platform for Simultaneous Monitoring of Spinal Cord Oxygenation, Perfusion and Metabolism

Hyperspectral imaging (HSI) enables spatial mapping of haemoglobin oxygenation and cytochrome-c-oxidase (CCO) redox changes in exposed neural tissue. In this study, we present a technical upgrade of an existing HSI platform by the integration of laser speckle contrast imaging (LSCI), enabling contemporaneous assessment of oxygenation, oxidative metabolism, relative perfusion and a derived relative oxygen metabolic index in exposed spinal cord in a rat preclinical model. The upgraded system (hNIR+) uses a shared optical path and a single camera to acquire HSI and LSCI images from the same field of view using a sequential acquisition workflow. Hyperspectral images are acquired at 11 discrete wavelengths (600, 630, 665, 784, 800, 818, 835, 851, 868, 881, and 894 nm) to estimate changes in oxygenated haemoglobin (HbO2), deoxygenated haemoglobin (HHb) and oxidized CCO (oxCCO). LSCI is implemented using 632.8-nm coherent illumination to generate relative blood-flow index (rBFI) maps. The LSCI subsystem was validated using a controlled blood-flow phantom, where the LSCI-derived inverse speckle contrast index showed a strong positive linear relationship with pump-controlled flow. The upgraded platform was then evaluated in proof-of-concept observations of the rat spinal cord under normoxia, hypoxia and recovery conditions. Representative multimodal maps and ROI-based time-course analysis showed clear dynamic responses in HbO2, HHb, oxCCO and rBFI during oxygen challenges that were consistent with the expected physiological responses of the spinal cord during hypoxia, hyperoxia and subsequent recoveries. Overall, these results demonstrate that integrating LSCI into HSI extends the platform to multimodal and simultaneous monitoring of haemodynamic-metabolic coupling.

physics.med-ph↗

Gold nanorod radioenhancement: interpreting Monte Carlo dose enhancement across geometry, composition, coating, and localization

Nanoparticle radiosensitization depends on interconnected NP properties, yet geometry, elemental composition, surface coating, and localization are typically studied separately, with their combined biological impact rarely assessed in one framework. Drawing on TOPAS Monte Carlo simulations from PhD research, we consolidated four investigations of metallic nanorods, mainly gold nanorods (AuNRs), examining these parameters across closely linked studies. The findings reveal that each parameter can influence over a distinct spatial range. Nanorod geometry primarily alters the spectrum and distribution of low-energy Auger-Meitner electrons emitted within tens of nanometers. Elemental composition induces significant near-field dose enhancement that rapidly attenuates and becomes negligible beyond roughly 1 um. Surface coatings reduce the fluence of secondary electrons below 3.5 keV, absorbed within about 150 nm of the surface, lowering physical dose contribution by 1-7%. Since coatings predominantly remove low-energy electrons, radiolytic yield changes through a shift in the electron energy spectrum, not directly following dose response. At the vascular scale, the enhanced dose region extends to approximately 10 um from the vessel wall, meaning NP localization determines where the biological target sits relative to radiation sources. Collectively, no single design parameter serves as a universal predictor of radiosensitization. Translating physical radioenhancement into biological effect depends on whether the target lies within each parameter's effective range. Furthermore, charged-particle equilibrium (CPE) correction, applied where beam confinement necessitated it, substantially alters calculated dose enhancement ratios. Consequently, scoring distance and CPE-corrected values should be explicitly reported to ensure comparability across studies.

physics.med-ph↗