arXiv · 2501.07180
Evaluating Robotic Approach Techniques for the Insertion of a Straight Instrument into a Vitreoretinal Surgery Trocar
Abstract
Advances in vitreoretinal robotic surgery enable precise techniques for gene therapies. This study evaluates three robotic approaches using the 7-DoF robotic arm for docking a micro-precise tool to a trocar: fully co-manipulated, hybrid co-manipulated/teleoperated, and hybrid with camera assistance. The fully co-manipulated approach was the fastest but had a 42% success rate. Hybrid methods showed higher success rates (91.6% and 100%) and completed tasks within 2 minutes. NASA Task Load Index (TLX) assessments indicated lower physical demand and effort for hybrid approaches.
Explore related subjects
Keep this discovery
Ross Henry, Martin Huber, Anestis Mablekos-Alexiou, Carlo Seneci, Mohamed Abdelaziz, Hans Natalius, Lyndon da Cruz, Christos Bergeles. 2025-01-13. Evaluating Robotic Approach Techniques for the Insertion of a Straight Instrument into a Vitreoretinal Surgery Trocar. https://arxiv.org/abs/2501.07180
Cite the original work for its findings. Save a collection to share your selection of sources.