Modeling and Measurement of Lead Tip Heating in Implanted Wires with Loops

Abstract

Purpose: To theoretically and experimentally study implant lead tip heating caused by radiofrequency (RF) power deposition in different wire configurations that contain loop(s). Methods: Maximum temperature rise caused by RF heating was measured at 1.5T on 20 insulated, capped wires with various loop and straight segment configurations. The experimental results were compared with predictions from the previously reported simple exponential and the adapted transmission line models, as well as with a long-wavelength approximation. Results: Both models effectively predicted the trends in lead tip temperature rise for all the wire configurations, with the adapted transmission line model showing superior accuracy. For superior/inferior (S/I)-oriented wires, increasing the number of loops decreased the overall heating. However, when wires were oriented right/left (R/L) where the x-component of the electric field is negligible, additional loops increased the overall heating. Conclusion: The simple exponential and the adapted transmission line models previously developed for, and tested on, straight wires require no additional terms or further modification to account for RF heating in a variety of loop configurations. These results extend the usefulness of the models to manage implanted device lead tip heating and provide theoretical insight regarding the role of loops and electrical lengths in managing RF safety of implanted devices.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Lydia J Bardwell Speltz, Seung-Kyun Lee, Yunhong Shu, Matt A Bernstein. 2024-12-19. Modeling and Measurement of Lead Tip Heating in Implanted Wires with Loops. https://arxiv.org/abs/2412.15120

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Lens-Aware Differentiable Beamforming for In Vivo Distributed Aberration Correction with Curvilinear Transducers

We previously introduced ultrasound autofocusing, an iterative model-based aberration correction technique that estimates local sound speed and incorporates it into beamforming to correct image distortion from heterogeneous media. In this work, we extend ultrasound autofocusing to curvilinear arrays and introduce advancements to the underlying model. A differentiable bent-ray tracing approach accounts for refraction through the transducer lens, while a new adaptive-grid initialization accounts for changes in speckle position with sound speed. The method is validated in silico and in calibrated sound speed phantoms. Our distributed aberration-correction method is then applied to a first large-scale in vivo evaluation comprising 313 liver acquisitions from 76 high-BMI human subjects. In images containing anechoic regions, contrast and CNR improved by $1.38 \pm 1.60$ dB (+18.0%) and $0.09 \pm 0.14$ (+10.2%), respectively. Improvements were also observed in speckle brightness (+20.3%), coherence factor (+13.1%), lag-one coherence (+2.7%), common-midpoint correlation coefficient (+0.7%), and common-midpoint phase error (-9.1%; lower is better), with all metric improvements statistically significant. Target structure and visibility also improved significantly. These results demonstrate the potential of ultrasound autofocusing for clinically applicable distributed aberration correction.

physics.med-ph

Closing the gap: A two-site MLC matching study for Varian TrueBeam Linear Accelerators

Purpose: To characterise and harmonise multileaf collimator (MLC) gap calibrations across four beam-matched TrueBeams and assess a common planning-system model. Methods: Four Millennium 120 TrueBeams were studied: TS1 (installed 2016), TS5 (2021), TS6 and TS7 (2024). Gap calibration was assessed by RayStation sweeping-gap measurement, feeler gauge, the EPID-based Stakitt Fence test, and a leaf-resolved Machine Performance Check (MPC). TS1 and TS5 were adjusted in service mode, TS5 also for MLC centreline offset. Ten clinical VMAT plans, and six acquired earlier on an unadjusted reference machine, were measured with an ArcCHECK 1220 against doses computed at 0.053 and 0.065 cm x-offsets in RayStation. Results: The 2024 machines had wider baseline gaps. All four methods registered the TS1 and TS5 changes in the same direction and within 0.060 mm of predictions. After adjustment TS1 and TS5 lay within the TS6/TS7 sweeping-gap and Stakitt ranges but read wider on the feeler gauge. A conventional picket-fence test on TS5 registered no change. Raising the RayStation x-offset parameter increased mean local 3%/2 mm pass rates from 92.3% to 94.8% on TS1 and 93.2% to 95.6% on TS5; all 24 plan-criterion pairs improved on the unadjusted machine. Conclusions: Mechanical, EPID, and sweeping gap MLC gap measurements responded consistently to the MLC adjustment, moving TS1 and TS5 towards the local reference range. Every global 3%/2 mm ArcCHECK measurement exceeded the TG-218 95% limit.

physics.med-ph

The Magnetic-Resonance-Based, Omni-Directional Hydrophone

This study aims to improve the directional sensitivity of magnetic resonance hydrophone (MRH) methods and allow for detection and quantification of waves traveling at any angle relative to the MRI bore-axis. Doing so may better integrate MRH technology into transcranial focused ultrasound neuromodulation therapies. Existing MRH implementations only encode acoustic waves propagating along the MRI bore-axis, precluding use with wide aperture transducers or temporal skull windows. This study expands the MRH technique into a MR-based Omni-directional Hydrophone (MRbOH) that can quantify the magnitude pressure and phase of waves propagating at any angle. The MRbOH method formulates a system of equations around uniquely sensitized images that, when solved, introduce sensitivity to oblique acoustic waves. The method was tested experimentally in a gel phantom that was insonated at $0^o$, $45^o$, and $90^o$ relative to the MRI bore-axis. The method reconstructed pressure maps at all three angles of propagation, including features such as peak pressure, partial reflections, and beamwidth. The MRbOH method measured an average beamwidth of 16.2 mm while the hydrophone measured a beamwidth of 14.5 mm. Average hydrophone-measured pressure within a region of interest was 416.19 $\boldsymbol{\pm}$ 78.07 kPa. Average MRbOH measured pressure in the same region at $0^o$, $45^o$, and $90^o$, respectively, were: 285.5 $\boldsymbol{\pm}$ 49.5 kPa, 193.13 $\boldsymbol{\pm}$ 43.55 kPa, 194.7 $\boldsymbol{\pm}$ 47.38 kPa. In addition, the study presents and experimentally verifies a statistical noise model that predicted the mean with $R^2 = 0.85$, and standard deviation with $R^2 = 0.83$.

physics.med-ph