arXiv · 2412.06260
Evaluation of Coincidence Time Resolution in a liquid xenon detector with silicon photomultipliers
Abstract
This work explores the combination of liquid xenon as a scintillating medium and silicon photomultipliers as a readout in Positron Emission Tomography (PET) for enhanced Time-Of-Flight resolution. We present the results of our first prototype optimized to maximize light collection using high-PDE, VUV-sensitive sensors and to minimize time fluctuations. We report a coincidence time resolution of 281 $\pm$ 2 ps FWHM, obtained using a $^{22}$Na calibration source. This result is competitive with the current state-of-the-art PET scanners and represents a significant step forward in the development of liquid xenon as a viable alternative to conventional scintillators in PET technology.
Explore related subjects
Keep this discovery
N. Salor-Iguiñiz, J. M. Benlloch-Rodríguez, R. Esteve, C. Romo-Luque, R. J. Aliaga, V. Álvarez, F. Ballester, R. Gadea, F. Monrabal, M. Querol, M. Rappaport, J. Rodríguez, J. Rodríguez-Ponce, S. Teruel-Pardo, J. F. Toledo, R. Torres-Curado, P. Ferrario, V. Herrero, J. J. Gómez-Cadenas. 2024-12-09. Evaluation of Coincidence Time Resolution in a liquid xenon detector with silicon photomultipliers. https://doi.org/10.1103/vy9h-twqf
Cite the original work for its findings. Save a collection to share your selection of sources.