Radon Removal Commissioning of the PandaX-4T Cryogenic Distillation System

Abstract

The PandaX-4T distillation system, designed for the removal of krypton and radon from xenon, is evaluated for its radon removal efficiency using a $^{222}$Rn source during the online distillation process. The PandaX-4T dark matter detector is employed to monitor the temporal evolution of radon activity. To determine the radon reduction factor, the experimental data of radon atoms introduced into and bypassed the distillation system is compared. The results indicate that the PandaX-4T distillation system achieves a radon reduction factor exceeding 190 at the flow rate of 10 slpm and the reflux ratio of 1.44. Gas-only online distillation process of a flow rate of 20 slpm is also conducted without observing significant reduction of radon levels in the detector. This observation suggests that the migration flow of radon atoms from the liquid phase to the gas phase is limited, and the flow rate of gas circulation and duration of the process are insignificant compared to the total xenon mass of 5.6 tons in the detector. This study provides the experimental data to support the efficient removal of radon at $\sim$Bq level using the PandaX-4T distillation system, which is the prerequisite of the radon background control in the detector. The further operation with higher flow rate will be applied for the upcoming science run in PandaX-4T.

Explore related subjects

Keep this discovery

BibTeXRIS

Xiangyi Cui, Zhou Wang, Jiafu Li, Shuaijie Li, Lin Si, Yonglin Ju, Wenbo Ma, Jianglai Liu, Li Zhao, Xiangdong Ji, Rui Yan, Haidong Sha, Peiyao Huang, Xiuli Wang, Huaxuan Liu. 2024-01-03. Radon Removal Commissioning of the PandaX-4T Cryogenic Distillation System. https://arxiv.org/abs/2401.01638

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det