Thermoluminescent characteristics of LiF:Mg,Cu,P and CaSO$_4$:Dy for low dose measurement

Abstract

Thermoluminescence (TL) characteristics for LiF:Mg,Cu,P, and CaSO$_4$:Dy under the homogeneous field of X-ray beams of diagnostic irradiation and its verification using thermoluminescence dosimetry is presented. The irradiation were performed utilizing a conventional X-ray equipment installed at the Hospital Juárez Norte of México. Different thermoluminescence characteristics of two material were studied, such as batch homogeneity, glow curve, linearity, detection threshold, reproducibility, relative sensitivity and fading. Materials were calibrated in terms of absorbed dose to the standard calibration distance and they were positioned in a generic phantom. The dose analysis, verification and comparison with the measurements obtained by the TLD-100 were performed. Results indicate that the dosimetric peak appears at 202$^o$C and 277.5$^o$C for LiF:Mg,Cu,P and CaSO$_4$:Dy, respectively. TL response as a function of X-ray dose showed a linearity behavior in the very low dose range for all materials. However, the TLD-100 is not accurate for measurements below $4mGy$. CaSO$_4$:Dy is 80\% more sensitive than TLD-100 and it show the lowest detection threshold, whereas LiF:Mg,Cu,P is 60\% more sensitive than TLD-100. All material showed very good repeatability. Fading for a period of one month at room temperature showed low fading LiF:Mg,Cu,P, medium and high for TLD-100 and CaSO$_4$:Dy. The results suggest that CaSO$_4$:Dy and LiF:Mg,Cu,P are suitable for measurements at low doses used in radiodiagnostic.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S. Del Sol Fernández, R. García-Salcedo, J. Guzmán Mendoza, D. Sánchez-Guzmán, G. Ramírez Rodríguez, E. Gaona, T. Rivera Montalvo. 2016-04-01. Thermoluminescent characteristics of LiF:Mg,Cu,P and CaSO$_4$:Dy for low dose measurement. https://doi.org/10.1016/j.apradiso.2016.02.011

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Enhanced micrometre spatial resolution and usable photon flux for hybrid pixel detectors via deep-learning-based X-ray localization

Micrometre spatial resolution is achieved for X-rays with the MÖNCH detector, a charge-integrating hybrid pixel detector with a 25 μm pixel pitch, by exploiting charge sharing between neighbouring pixels. The conventional analytical interpolation method, however, relies on a global mapping derived from the charge-sharing statistics and is applicable only to isolated single-photon clusters, which limits both the achievable resolution and the usable photon flux. To overcome these limitations, we trained deep learning models to localize single-photon and pile-up events based on high-fidelity simulation data. For knife-edge measurements at 12 keV and a low occupancy of 0.74% (photons per pixel and frame), where single-photon events dominate, the proposed method reaches a of the edge spread function of 1.36 μm, compared with 1.54 μm for the conventional eta interpolation. We demonstrate experimentally that resolving up to three photons per cluster raises the reconstruction efficiency from 59% to 93% at a 2.4{\times} higher occupancy of 1.81%, while preserving a of 1.39 μm.

physics.ins-det↗

From Tunnel Overburden Reconstruction to Tidal Measurements via atmospheric muon

Muography with scintillator-based detectors enables the investigation of density distributions within large structures. The Hankuk Atmospheric-muon Wide Landscaping (HAWL) project is the first real-time portable muon tomography. It successfully charted the mountainous region of eastern Korea by measuring cosmic-ray muons with a detector operated in a moving vehicle. HAWL achieved a tunnel length accuracy of 6.0 % and an overburden measurement range of 8-400 meter-water-equivalent (m.w.e.). The HAWL upgrade (HAWLU) features a more compact design and enables trajectory reconstruction. Using this upgraded system, we measured tidal water-level variations with meter-scale precision over 31.6 days in the Boryeong Undersea Tunnel. The measured muon rate decreased by $1.72 \pm 0.04$ % per meter of increasing tidal level, while the reconstruction of undersea tunnel scan independently validated the detector performance. These results demonstrate the capabilities of dynamic muography in two approaches, one with a portable detector in motion and the other with a detector measuring a target that changes over time.

physics.ins-det↗

Autocorrelation-based beam shape optimization in scanning electron microscopy

Many automatically focusing algorithms for scanning electron microscopes exist. Typically, these optimize a single derived metric (such as image sharpness) without considering the intrinsic beam shape. We describe an algorithm to automatically focus and stigmate scanning electron microscopes based on the autocorrelation of captured images. In the first part of this work, we develop the theoretical basis of using the autocorrelation to approximate the current beam shape. In the second part, we use this to condition the beam shape on various surfaces, using Bayesian optimization to tune both working distance and, in particular, stigmation. Through a comparison of the full width at half maximum of the autocorrelation arrays before and after optimization and a visual comparison of images taken of the sample surface before and after optimization, we show that the algorithm is able to successfully optimize the scanning parameters. Overall, our algorithm can effectively focus on a variety of topographies at different magnifications.

physics.ins-det↗