arXiv · physics/0610174
A stiffness sensor to help in the diagnosis and the surgery of orbital pathologies
Abstract
Proptosis is characterized by a protrusion of the eyeball due to an increase of the orbital tissue volume. To recover a normal eyeball positioning, the most frequent surgical technique (BROD technique) consists in the osteotomy of orbital walls combined with a loading on the eyeball to initiate tissue decompression. In this paper, a stiffness sensor device is proposed to (1) provide to the surgeon pre, intra and post-operative data concerning the stiffness of the intra-orbital soft tissues, and (2) provide constitutive parameters to the Finite Element model of the intra-orbital tissues already developed by the authors and used to predict consequences orbital surgery.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Vincent Luboz, Dominique Ambard, Franck Boutault, Pascal Swider, Yohan Payan. 2006-10-22. A stiffness sensor to help in the diagnosis and the surgery of orbital pathologies. https://arxiv.org/abs/physics/0610174
Cite the original work for its findings. Save a collection to share your selection of sources.