Development and operation of a pixel segmented liquid-filled linear array for radiotherapy quality assurance

Abstract

A liquid isooctane (C$_{8}$H$_{18}$) filled ionization linear array for radiotherapy quality assurance has been designed, built and tested. The detector consists of 128 pixels, each of them with an area of 1.7 mm $\times$ 1.7 mm and a gap of 0.5 mm. The small pixel size makes the detector ideal for high gradient beam profiles like those present in Intensity Modulated Radiation Therapy (IMRT) and radiosurgery. As read-out electronics we use the X-Ray Data Acquisition System (XDAS) with the Xchip developed by the CCLRC. Studies concerning the collection efficiency dependence on the polarization voltage and on the dose rate have been made in order to optimize the device operation. In the first tests we have studied dose rate and energy dependences, and signal reproducibility. Dose rate dependence was found lower than 2.5 % up to 5 Gy min$^{-1}$, and energy dependence lower than 2.1 % up to 20 cm depth in solid water. Output factors and penumbras for several rectangular fields have been measured with the linear array and were compared with the results obtained with a 0.125 cm$^{3}$ air ionization chamber and radiographic film, respectively. Finally, we have acquired profiles for an IMRT field and for a virtual wedge. These profiles have also been compared with radiographic film measurements. All the comparisons show a good correspondence. Signal reproducibility was within a 2% during the test period (around three months). The device has proved its capability to verify on-line therapy beams with good spatial resolution and signal to noise ratio.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J. Pardo, L. Franco, F. Gomez, A. Iglesias, A. Pazos, J. Pena, R. Lobato, J. Mosquera, M. Pombar, J. Sendon. 2004-12-01. Development and operation of a pixel segmented liquid-filled linear array for radiotherapy quality assurance. https://doi.org/10.1088/0031-9155%2F50%2F8%2F006

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Probabilistic Modeling of Photon Interaction Chains in a Silicon Detector

Compton camera imaging relies on the positions and energies of multiple photon interactions recorded in the detector. In silicon detectors that resolve individual interactions, these measurements do not directly provide the causal interaction order, and in nuclear medicine applications the photon may also have scattered in the object before entering the detector. We present a probabilistic model for photon interaction chains in a silicon detector and use it for two inference tasks at the event level: interaction order inference and classification of primary versus object-scattered photons. The model combines photon transport based on Compton scattering and attenuation with a detector response model. It is evaluated on Monte Carlo data for monoenergetic point sources at 140~keV and 511~keV, with and without a cylindrical water phantom. For interaction chains of length two to four, the maximum a posteriori order estimate gives high accuracy in the case without a phantom, exceeding 0.95 for all chain lengths at 511~keV, and decreasing with increasing phantom radius. The gap between the two highest posterior order probabilities provides a confidence measure for the ordering estimate and can be used to increase accuracy at the cost of reduced retention. For scatter classification, likelihood-ratio testing yields receiver operating characteristic curves with area under the curve above 0.85 across all phantom radii considered. These results support the use of probabilistic modeling of measured interaction sequences for inference directly at the event level, and show that both interaction ordering and scatter discrimination can be treated within a common framework.

physics.med-ph↗

Perioperative shear wave elastography for the evaluation of spinal cord release in pediatric dysraphism: first experience in 21 patients

OBJECTIVE Shear wave elastography (SWE) is a recent method for evaluating soft tissue stiffness. Its application to the spinal cord is limited by the ultrasound bone barrier. Spinal dysraphism can cause neurological symptoms impairing bladder, bowel, and lower limb function due to a mechanical stress known as tethered cord syndrome. Detethering of the spinal cord is a part of surgical treatment. This study aimed to assess the feasibility of intraoperative elastography to guide surgery in patients with spinal dysraphism involving traction of the spinal cord. METHODS This single-center retrospective study included all patients who underwent surgery for spinal dysraphism between November 2023 and July 2024. SWE was performed under general anesthesia with the patient in the prone position before and after spinal cord detethering. All SWE measurements were performed using the Aixplorer system. Demographic, clinical, and radiological data (age at surgery, neurological status, and MRI findings) were collected. RESULTS Twenty-one patients with a mean $\pm$ SD (range) age of 3.7 $\pm$ 5.3 (0--17.5) years were included. Etiologies included open saccular dysraphism (n = 3), spinal cord conus lipoma (n = 6), limited dorsal myeloschisis (n = 3), and fibrolipomatous filum (n = 9). In 85% of patients, at least 1 clip was of sufficient quality for analysis both before and after release. The mean $\pm$ SD (range) spinal cord stiffness at the beginning of surgery was 25.2 $\pm$ 9.2 (12.3--39.1) kPa and 11.2 $\pm$ 3.8 (6.2--16.5) kPa at the end of release, with a significant stiffness decrease (p = 0.018). No complications attributable to SWE were observed. CONCLUSIONS SWE of the spinal cord in patients with spinal dysraphism is a safe and feasible imaging method. This technique may assist in assessing the adequacy of untethering during surgery. However, no established normative SWE values for the pediatric spinal cord currently exist. Until it does, the intraoperative interpretability and clinical utility of this technique remain limited.

physics.med-ph↗

Influence of Geometry and Tissue Damage on the Mechanical Behaviour of Skin Meshing

Skin meshing increases graft coverage, yet the clinically achieved expansion often falls below the manufacturer's nominal meshing ratio. Previous elastic analyses have demonstrated that mesh topology and skin anisotropy strongly influence graft expansion; however, the effect of progressive damage at larger deformation remains unexplored. This study therefore investigates how progressive damage affects the expansion and mechanical response of skin meshes. The performance of auxetic mesh geometries is compared with conventional Zimmer mesher (ZM) geometry. Human back skin was modelled using a one-term Ogden ground matrix and a Gasser Ogden Holzapfel (GOH) collagen network, with separate scalar Kachanov-type continuum damage variables for each constituent. The auxetic geometries, including 1st- and 2nd-order alternative slits (AS1 and AS2), I-shaped (IS) and H-shaped (HS) re-entrant patterns, and ZM, were analysed with Langer's lines oriented either parallel or perpendicular to the loading direction. Hyperelastic and damage-based formulations were compared up to 25% strain, and the damage-based simulations were subsequently extended to 40% strain. ZM was further loaded to 155% strain; the area ratios attained by HS at 40% strain were matched by ZM at 142.1% and 124.7% strain for Langer's lines parallel and perpendicular to loading, respectively. Results indicate that HS achieves greater coverage than ZM for a given stretch, whereas its damage advantage depends on the orientation of Langer's lines.

physics.med-ph↗