arXiv · 2611.14040
Characterization of BNLs 20 and 30 micrometer thick LGADs for the Electron Ion Collider
Abstract
The 20 and 30 micrometer thick Low-Gain Avalanche Detectors (LGADs) produced at Brookhaven National Laboratory (BNL) were evaluated. Detailed electrical measurements indicate well-formed and stable gain layers with consistently low leakage currents. Transient Current Technique (TCT) was used to evaluate the gain. Fast-laser timing studies demonstrate time resolutions at the level of a few tens of picoseconds. To assess operational longevity under realistic conditions for Electron-Ion Collider sub-detector systems, devices were irradiated with 30 MeV and 60 MeV protons to fluences of 10^13 to 10^14 neq/cm^2, enabling a focused study of radiation-induced effects on gain stability and timing performance.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
F. Zareef, J. J. Chwastowski, M. Pycior, I. Ahmad, W. Fryt, G. Giacomini, S. Kusyk, A. Obłąkowska-Mucha, P. Płatek, T. Polański, P. Rzeźnik, J. Swakoń, T. Szumlak, A. Tricoli, D. Wróbel. 2026-11-10. Characterization of BNLs 20 and 30 micrometer thick LGADs for the Electron Ion Collider. https://arxiv.science/abs/2611.14040
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite arXiv Science by Indiciae: BibTeX · RIS.