PhysiCal-PET: Physics-driven neural networks for the rapid position Calibration of PET Scanners

Abstract

Depth-of-interaction (DOI) information is essential for mitigating parallax error and preserving uniform spatial resolution in high-resolution and long-axial-field-of-view positron emission tomography (PET). Conventional DOI calibration relies on mechanically scanned collimated sources or detector-specific bench measurements, which become impractical as the number of detector modules grows. We present PhysiCal-PET, a physics-driven neural-network framework for fast, label-free DOI calibration of fully assembled PET scanners. Coincidences are acquired from an uncollimated point source placed at a few known positions, and a network mapping DOI-sensitive detector signals to interaction depth is trained by minimizing the distance between the known source position and the line of response (LOR) defined by the two predicted interaction positions, so that the annihilation geometry alone provides supervision, without measured DOI labels. In realistic simulations, the DOI mean absolute error decreased in parallel with the geometric loss and converged to 0.98 mm, confirming that the loss is a faithful surrogate for the unobservable DOI error. Applied to a preclinical scanner with 144 pixelated dual-ended-readout detector modules, the method calibrated the entire system from 11 manually placed source positions in 55 min of acquisition, reducing the trimmed source-to-LOR distance by 24%, improving the radial point-source FWHM from 4.99 to 2.90 mm (42%), and raising the Derenzo peak-to-valley ratio of the 1.7 mm rods from 1.28 to 1.88 while separating the 1.2 mm rods, which remain merged without calibration. Requiring no collimation, motion stages, or reference measurements, and scaling favorably with system sensitivity, PhysiCal-PET offers a practical route toward autonomous, routine DOI calibration of fully assembled PET scanners with many detectors.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Mojahed Abushawish, Pablo Galve, Iván Colmenarejo Allavena, Carlos Galindo Morata, Sara Gaitan Dominguez, Jose Manuel Udias. 2026-10-16. PhysiCal-PET: Physics-driven neural networks for the rapid position Calibration of PET Scanners. https://arxiv.org/abs/2610.24193

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Implementing the EIT complete electrode model through Robin boundary conditions

Objective. Medical electrical impedance tomography systems prescribe electrode currents. Accurate voltages, even on passive electrodes, require the complete electrode model (CEM), the most accurate model known for skin electrodes. Its current constraints specify total electrode currents, whereas standard finite element boundary interfaces impose local conditions. We describe how to implement the CEM through Robin boundary conditions, which are available in standard FEM packages. Approach. Prescribed electrode potentials give Robin problems whose unit-potential responses define an electrode transadmittance matrix. We prove equivalence to the conventional augmented CEM and formalise the discrete energy, equivalence and reference arguments in Lean 4. Numerical comparisons use two-dimensional P1 finite elements in DOLFINx, FreeFEM, GetFEM, MFEM and NGSolve, with the same nodal reference potential as EIDORS. Main results. For identical assembled blocks, the two procedures have the same solution set whenever the Robin matrix is invertible. Across the five packages in dimensionless test cases, maximum absolute electrode-potential differences from EIDORS were below $5\times10^{-13}$ (2692 nodes), and errors against an exact rational-arithmetic reference were below $1.1\times10^{-13}$ (2535 nodes). Significance. Standard Robin solves and consistent electrode-current integration provide a portable route to the current-driven CEM, with current constraints enforced through a small electrode system.

physics.med-ph↗

Attenuation corrected activity reconstruction for listmode TOF-PET data

Time-of-flight (TOF) positron emission tomography (PET) emission data allow joint estimation of the activity and attenuation images. The methods proposed so far rely on reconstructions from TOF histograms of the measurements. However, with the rapid data increase of total-body (TB) scanners and the ever improving TOF and intrinsic resolutions, more memory and computation efficient techniques for joint reconstruction need to be developed. In this work, a modified version of the maximum likelihood activity and attenuation correction factors algorithm (MLACF) is presented, that jointly estimates the activity image together with the attenuation factors (AFs) from listmode emission data. The method does not require TOF histogramming of the data, however it still relies on a coarsely sampled (i.e. mashed) non-TOF histogram of the emission data. The activity distribution is subsequently quantified by reconstructing and enforcing known expected tissue attenuation on an attenuation reconstruction from the estimated AF values. The algorithm is evaluated using 3D simulations as well as clinical TOF-PET measurements of a patient scan and compared to reference reconstructions. As could be expected, we find the reconstruction accuracies improve with an "optimal" histogramming of the TOF emission data, increasing the signal to noise ratio of the data used for joint estimation. With an "optimal" histogramming of the measurements, the attenuation corrected activity distributions from TOF emission data only are comparable to the gold-standard in PET/CT.

physics.med-ph↗

Localizing and quantifying cerebral arterial stenosis from sparse velocity waveforms using a conditional neural operator

Cerebral arterial stenosis alters blood flow throughout the vascular network, complicating the estimation of its location and severity from sparse velocity observations. We developed a conditional neural operator that predicts cerebral hemodynamics for candidate stenoses and enables inverse estimation through waveform comparison. Training data were generated using a sequential 0D-1D model that simulates pressure, cross-sectional area, and velocity under varying anatomical, physiological, and stenosis conditions. The operator encodes a single stenosis within vessel tokens and uses global attention to represent its effects across the network. Severity-gated coupling connects these representations to a shared field decoder, while paired healthy simulations constrain baseline predictions. After training, model weights remain fixed, and stenosis location and severity are estimated by minimizing the mismatch between predicted and observed midpoint velocity waveforms. Network-averaged relative validation errors were 0.14% for pressure, 0.59% for area, and 2.79% for velocity. For an eight-vessel observation configuration, localization accuracy across 20 candidate vessels was 99.4% without added noise and 96.0% at the 30% noise setting. Comparisons between configurations with equal numbers of observed vessels showed that measurement placement influenced recovery. Errors and localization ambiguities were greater in selected communicating and distal arteries. These results demonstrate the feasibility of surrogate-based stenosis estimation in simulated cerebral networks and highlight the importance of observation placement. Evaluation was limited to unseen stenosis scenarios within sampled anatomical and physiological conditions. Generalization to new individuals and clinical measurements requires further investigation.

physics.med-ph↗