arXiv · 2609.21618
A compact fibre-coupled active beamstop for fast time-resolved monitoring of synchrotron X-ray beams
Abstract
We have developed a compact fibre-coupled active beamstop for simultaneous interception and time-resolved monitoring of the direct X-ray beam in synchrotron diffraction experiments. The complete beamstop tip has a maximum transverse dimension of 1 mm and incorporates a 700 $μ$m-wide Ce:YAG scintillator. The scintillator converts the intercepted X-rays into visible light, which is transported through a multimode optical fibre to a silicon photodiode at the bottom of the blade that is connected to a dedicated fast amplifier electronics. Tests with a pulsed laser yielded a system decay time of 69.4 ns, consistent with reported luminescence decay times for Ce:YAG. Measurements at the P14.EH2 end station of PETRA III demonstrated the detection of individual bunch signals separated by 192 ns in the 40-bunch operating mode. The active beamstop also provided an in-situ timing signal for synchronising an experimental sequence with the approximately 190 $μ$s transmission window of a mechanical X-ray chopper. The device therefore combines a compact obstruction of the direct beam with fast monitoring of X-ray timing and flux during diffraction data acquisition.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Friedjof Tellkamp, David von Stetten, Martin Kollewe. 2026-09-18. A compact fibre-coupled active beamstop for fast time-resolved monitoring of synchrotron X-ray beams. https://arxiv.org/abs/2609.21618