Physical design of cold neutron direct geometry inelastic spectrometer at China Spallation Neutron Source
Abstract
The Cold-Neutron Inelastic Spectrometer (CNIS) is a direct-geometry, time-of-flight instrument designed for China Spallation Neutron Source (CSNS) and optimized to probe low-energy lattice and magnetic excitations. The instrument integrates a long flight path with bent supermirror guides and an elliptical-focusing geometry to suppress high-energy background while improving cold-neutron delivery to the sample. A flexible multi-disk chopper suite provides pulse shaping, band selection and monochromatization, enabling multi-$E_\textrm{i}$ operation. Modular features, including an interchangeable high-focusing guide insert, radial collimation and a vacuum ``airbox'' for simplified sample-environment integration, enhance signal-to-noise and operational versatility. Through combined flight-path and chopper optimization, CNIS achieves excellent routine-mode energy resolution and can reach approximately $\sim 1\%$ in a dedicated high-resolution configuration. CNIS is planned to commence user operation in 2029, offering a highly flexible platform for cold-neutron inelastic scattering studies.
Explore related subjects
Keep this discovery
Qian Zhao, Xiaowen Zhang, Songwen Xiao, Wei Luo, Zecong Qin, Zhuang Xu, Yu Feng, Xin Tong. 2026-05-11. Physical design of cold neutron direct geometry inelastic spectrometer at China Spallation Neutron Source. https://doi.org/10.1063/5.0321019
Cite the original work for its findings. Save a collection to share your selection of sources.