arXiv · 2603.12010
First in-vivo human magnetic particle imaging
Abstract
Magnetic particle imaging (MPI) is a tracer-based technique that directly detects the distribution of magnetic iron-oxide nanoparticles with millisecond temporal resolution and no tissue background. Despite extensive preclinical work, in-vivo application of MPI in humans has not previously been reported. Here, we report the first in-vivo human MPI angiography, visualizing venous perfusion of the upper extremity using a human-scale scanner and clinically approved ferucarbotran. Under identical procedural conditions, we performed X-ray digital subtraction angiography as the clinical gold standard. MPI visualized major superficial and deep veins, including inflow, branching, valve filling, and clearance dynamics in real time with 2 frames per second. These results establish magnetic particle imaging as a clinically translatable modality for radiation-free vascular imaging in humans and mark the transition of MPI from preclinical research to first clinical application.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Patrick Vogel, Thomas Kampf, Martin A. Rückert, Johanna Günther, Teresa Reichl, Thorsten A. Bley, Volker C. Behr, Philipp Gruschwitz, Viktor Hartung. 2026-03-12. First in-vivo human magnetic particle imaging. https://arxiv.org/abs/2603.12010
Cite the original work for its findings. Save a collection to share your selection of sources.