An image-guided high-precision research platform for ultra-high dose rate spinal cord toxicity studies

Abstract

Objective: While FLASH radiotherapy is recognized for short-term normal tissue sparing, its durability in late-responding organs remains uncertain, limiting clinical adoption. With its clinical importance and steep dose-response, the spinal cord is an ideal model for evaluating FLASH effect on late toxicity. This work introduces a robust image-guided research platform for high-precision irradiation at both CONV and UHDR to enable FLASH late toxicity studies using a rat spinal cord model. Approach: A modified LINAC was employed to irradiate the C1-T2 rat spinal cord with 18 MeV UHDR and CONV beams. A custom rat immobilization device, a portable X-ray imaging system, and an ion-chamber-based UHDR output monitoring system were integrated to ensure accurate C1-T2 localization and precise dose delivery. A Monte Carlo (MC) dose engine was developed to provide accurate dosimetry and support interpretation of in vivo results. Scintillator measurements were performed within the spinal cord to verify MC results and the precision of our platform. Results: We achieved submillimeter C1-T2 setup accuracy and maintained submillimeter intrafraction motion. Ion chamber readings showed linear correlation with UHDR output. MC indicated uniform irradiation along the central ~13 mm cord. Our CONV beam exhibited distribution close to that of the UHDR beam, with difference <3%, isolating dose rate as the only variable. Scintillator-measured dose agreed with MC within 4%, confirming both MC accuracy and the platform's high-precision delivery. Significance: We developed the first comprehensive, image-guided preclinical platform for accurate UHDR and CONV irradiation to investigate FLASH-mitigated spinal cord toxicity in rats. This work thus establishes a robust foundation for systematic evaluation of the FLASH effect in late-responding organs and for determining clinical applicability of FLASH.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Banghao Zhou, Lixiang Guo, Yi-Chun Tsai, Albert van der Kogel, John Wong, Iulian Iordachita, Kai Jiang, Weiguo Lu, Paul Medin, Ken Kang-Hsin Wang. 2025-11-12. An image-guided high-precision research platform for ultra-high dose rate spinal cord toxicity studies. https://arxiv.org/abs/2511.08974

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Fail-closed conformal multiresolution error control for conservative three-dimensional dose remapping in synthetic phantoms

Conservative dose remapping can be formulated by transporting mass and deposited energy across nonmatching grids, but fixed high-sampling calculations use the same work for easy and difficult cases. We developed conformal multiresolution error control (CoMERC), a fail-closed multilevel quasi-Monte Carlo (QMC) controller for reference-relative target-cell allocation error in a fixed piecewise-constant source model with known deformation. Two randomized nested replicates were evaluated at a probe level of 4 and candidate levels of 8 and 16. Four endpoints were controlled jointly: reference-mass-weighted global, high-gradient, and density-gradient-threshold root-mean-square error, plus eligible-cell maximum absolute error. A split-conformal multiplier was calibrated from 240 synthetic cases, frozen, and evaluated on 400 independent cases from the same generator and 120 cases from a prespecified shifted generator. In the primary set, joint coverage was 394/400 (98.50%; one-sided 95% lower limit, 97.06%), 388/400 cases (97.00%; lower limit, 95.18%) received an output, and 0/388 released outputs exceeded any tolerance (one-sided 95% upper limit, 0.769%). Final actions were level 8 for 190 cases, level 16 for 198, and ABSTAIN for 12. The frozen policy implied mean relative production-sample work of 0.5844 compared with always running both randomized replicates through level 16; the prespecified 95th-percentile bootstrap upper limit was 0.6194. All seven primary gates passed. In the shifted-generator set, 113/120 cases were released and one released output exceeded a tolerance. CoMERC provided marginal finite-sample error control and lower modeled production-sample work in the locked synthetic population, but not patient-level, registration-level, conditional-on-release, or clinical safety validation.

physics.med-ph

Dosimetric equivalence of deep learning prostate contours after LDR brachytherapy: pre-declared margins, patient-level acceptance thresholds and the incremental predictive value of DVH indices

Background and purpose: Dose Volume Histogram (DVH) indices remain the main dose-effect metrics for toxicity prediction, but they depend on how contours were made. Inter-observer variability (IOV) is unavoidable and clinically accepted, so the question for automatic segmentation addresses equivalence: do automatic contours produce DVH errors comparable to human IOV, and can these indices predict patient-reported toxicity? Materials and methods: In 429 patients treated with iodine-125 LDR-BT monotherapy, indices from expert manual delineation were compared with a deterministic and a Bayesian nnUNet on a fixed dose distribution, by two-one-sided tests against margins set from CT contouring IOV. Logistic regression gave patient-level thresholds at 90\,\% probability of equivalence. In 380 patients, eleven DHV indices were added to a clinical baseline predicting change in International Prostate Symptom Score (IPSS) at six horizons over 5 years, with nested cross-validation, bootstrap intervals and corrected $t$-tests. Results: All cohort-level comparisons of DVH indices were declared equivalent, with no interval consuming more than 44\,\% of its equivalence margin. Individual agreement was weaker with equivalence rates from 55.9\,\% to 89.7\,\%, depending on the DVH index and automatic segmentation model. Thresholds ranged from 0.864 to 0.958 Dice. No DVH block improved IPSS prediction at any horizon. The largest improvement declared by bootstrap intervals was 0.12 IPSS points, well below the minimal clinically important difference, across all learners. Conclusion: Automatic contours matched expert dosimetry within human IOV on average, but individual equivalence requires a volume dependent quality metric threshold definition. Our DVH indices panel provides no significant predictive power for IPSS, irrespective of segmentation source and horizon.

physics.med-ph

The Electrodynamic Basis of Dichroism-Mediated Polarization Perception

Humans see the polarization of light through entoptic percepts arising from the macula's Henle fiber layer, where xanthophyll pigments absorb preferentially across the radiating fibers. From the layer's complex dielectric tensor alone, Maxwell's equations in Berreman $4\times4$ form yield its Mueller matrix, set by four scalars $\{A,B,C,D\}$. Its intensity channel depends only on the dichroic pair $A$ and $B$, which fix the percepts at a maximum contrast $|B|/A\approx0.05$. One relation generates the whole dichroism-mediated family: Haidinger's brushes under a uniform field, their dark arms perpendicular to the $\mathbf{E}$-vector; fractured brushes under spatially varying fields; and $N$-fold brushes under vector-vortex illumination.

physics.med-ph