Improved energy-based scatter estimation by incorporating local energy spectra and accelerating the parametric fitting

Abstract

PET requires accurate, precise, and efficient scatter correction techniques. Conventional scatter estimation typically relies on tail-fitted single-scatter simulation (SSS) strategy. However, the accuracy of tail-fitted SSS is limited, for example, by mismatches between the attenuation image and the PET emission data or by the presence of activity outside the FOV. These shortcomings can be addressed using energy-based scatter estimation (EBSE), as recently proposed by Efthimiou et al. and Hamill et al. The aim of this work is to 1. improve the accuracy of EBSE by accounting for the LOR dependence of the energy spectrum of unscattered photons, 2. improve the computational speed of EBSE through better initialization and a more efficient optimization algorithm. The proposed improved EBSE method models the energy probability density function (PDF) of both single and multiple scattered photons, and incorporates a position-dependent energy PDF for unscattered photons. These energy PDFs form the basis of two forward models used for scatter estimation based on 2D energy histograms. The performance of these models were evaluated using GATE Monte Carlo simulations and a NEMA phantom acquisition on a GE SIGNA PET/MR scanner. Furthermore, we assessed the stability of EBSE across the forward models by varying the number of counts in the 2D energy histograms via data mashing. EBSE outperformed tail-fitted SSS, particularly in regions near out-of-FOV activity. Our GATE simulations showed that incorporating a local energy for unscattered photons improves off-center regional quantification by approximately 2% points. Additionally, improved initialization combined with the NEGML optimizer enabling execution on a mashed TOF sinogram in 12 minutes on six-core CPU. The proposed method enhances both the accuracy and computational efficiency of EBSE, making it well-suited for clinical applications.

Explore related subjects

Keep this discovery

BibTeXRIS

Seyed Amir Zaman Pour, Ahmadreza Rezaei, Floris Jansen, Kristof Baete, Georg Schramm, Johan Nuyts. 2025-04-22. Improved energy-based scatter estimation by incorporating local energy spectra and accelerating the parametric fitting. https://arxiv.org/abs/2504.15969

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph