arXiv · 2409.09531
Direct frequency comb cavity ring-down spectroscopy: Enhancing Sensitivity and Precision
Abstract
We present a novel approach to cavity ring-down spectroscopy utilizing an optical frequency comb as the direct probe of the Fabry-Perot cavity, coupled with a time-resolved Fourier transform spectrometer for parallel retrieval of ring-down events. Our method achieves high spectral resolution over a broad range, enabling precision measurements of cavity losses and absorption lineshapes with enhanced sensitivity. A critical advancement involves a stabilization technique ensuring complete extinction of comb light without compromising cavity stabilization. We demonstrate the capabilities of our system through precision spectroscopy of carbon monoxide rovibrational transitions perturbed by argon.
Explore related subjects
Keep this discovery
Romain Dubroeucq, Dominik Charczun, Piotr Masłowski, Lucile Rutkowski. 2024-09-14. Direct frequency comb cavity ring-down spectroscopy: Enhancing Sensitivity and Precision. https://arxiv.org/abs/2409.09531
Cite the original work for its findings. Save a collection to share your selection of sources.