arXiv · 2401.04488
Development and validation of a cryogenic far-infrared diffraction grating spectrometer used to post-disperse the output from a Fourier transform spectrometer
Abstract
Recent advances in far-infrared detector technology have led to increases in raw sensitivity of more than an order of magnitude over previous state-of-the-art detectors. With such sensitivity, photon noise becomes the dominant noise component, even when using cryogenically cooled optics, unless a method of restricting the spectral bandpass is employed. The leading instrument concept features reflecting diffraction gratings which post-disperse the light that has been modulated by a polarizing Fourier transform spectrometer (FTS) onto a detector array, thereby reducing the photon noise on each detector. This paper discusses the development of a cryogenic (4 K) diffraction grating spectrometer which operates over the wavelength range from 285 - 500 $\mu$m and was used to post-disperse the output from a room-temperature polarizing FTS. Measurements of the grating spectral response and diffraction efficiency are presented as a function of both wavelength and polarization to characterize the instrumental performance.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Alicia M. Anderson, David A. Naylor, Brad G. Gom, Matthew A. Buchan, Adam J. Christiansen, Ian T. Veenendaal. 2024-01-09. Development and validation of a cryogenic far-infrared diffraction grating spectrometer used to post-disperse the output from a Fourier transform spectrometer. https://doi.org/10.1063/5.0177603
Cite the original work for its findings. Save a collection to share your selection of sources.