Calculation of true coincidence summing correction factor for a Broad Energy Germanium (BEGe) detector using standard and fabricated sources
Abstract
The true coincidence summing (TCS) correction factor for a Broad Energy Germanium (BEGe) detector has been calculated at far and close geometry measurement using multi-energetic radioactive $\gamma$-ray sources $^{60}$Co, $^{133}$Ba and $^{152}$Eu. The correction factors were calculated using experimental method and analytical method. Photopeak efficiency and total efficiency required to calculate the correction factor were obtained using Geant4 Monte Carlo simulation code. A few standard as well as fabricated mono-energetic sources were also included in the $\gamma$-ray efficiency measurements. The simulated efficiencies of mono-energetic $\gamma$-ray sources were matched to experimental $\gamma$-ray efficiencies by optimizing the detector parameters. The same parameters were used to obtain the photopeak and total efficiency for $\gamma$-ray of our interest and coincident $\gamma$-ray. Analytical correction factors and experimental correction factors were found in good agreement with each other.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Ashish Gupta, M. Shareef, Munmun Twisha, Saikat Bhattacharjee, Gopal Mukherjee, Satya Samiran Nayak, Sansaptak Basu, S. Dasgupta, J. Datta, S. Bhattacharyya, A. Mukherjee. 2023-02-06. Calculation of true coincidence summing correction factor for a Broad Energy Germanium (BEGe) detector using standard and fabricated sources. https://doi.org/10.1016/j.apradiso.2023.110966