arXiv · 2203.04896
A backing detector for order-keV neutrons
Abstract
We have designed and tested a large-area (0.15~m$^2$) neutron detector based on neutron capture on \ce{^{6}Li}. The neutron detector design has been optimized for the purpose of tagging the scattering angle of keV-scale neutrons. These neutron detectors would be employed to calibrate the low-energy ($<$100 eV) nuclear recoil in detectors for dark matter and coherent elastic neutrino nucleus scattering (CE$\nu$NS). We describe the design, construction, and characterization of a prototype. The prototype is designed to have a tagging efficiency of $\sim$25\% at the relevant $\mathcal{O}$(keV) neutron energies, and with a mean capture time of $\sim$17$~\mu$s. The prototype was characterized using a \ce{^{252}Cf} neutron source and agreement with the simulation was observed within a few percent level.
Explore related subjects
Keep this discovery
A. Biekert, L. Chaplinsky, C. W. Fink, M. Garcia-Sciveres, W. C. Gillis, W. Guo, S. A. Hertel, G. Heuermann, X. Li, J. Lin, R. Mahapatra, D. N. McKinsey, P. K. Patel, B. Penning, H. D. Pinckney, M. Platt, M. Pyle, R. K. Romani, A. Serafin, R. J. Smith, B. Suerfu, V. Velan, G. Wang, Y. Wang, S. L. Watkins, M. R. William. 2022-03-09. A backing detector for order-keV neutrons. https://doi.org/10.1016/j.nima.2022.166981
Cite the original work for its findings. Save a collection to share your selection of sources.