arXiv · 2202.08746
Detection of low-energy X-rays with 1/2 and 1 inch LaBr3:Ce crystals read by SIPM arrays
Abstract
LaBr3:Ce crystals, with photomultiplier or single SiPM readout (up to 3x3 mm$^2$) have been introduced for radiation imaging in medical physics. An R&D was pursued with 1/2" and 1" LaBr3:Ce crystals, from different producers, to realize compact large area detectors (up to some cm $^2$ area) with SiPM array readout, aiming at high light yields, good energy resolution,good detector linearity and fast time response for low-energy X-rays. The study was triggered by the FAMU experiment at the RIKEN-RAL muon facility, aiming at a precise measurement of the proton Zemach radius to solve the so-called "proton radius puzzle". The goal is the detection of characteristic X-rays around 130 KeV. Other applications may be foreseen for homeland security and $\gamma$-ray astronomy. Results were obtained with a direct readout based on a CAEN V1730 FADC, with no need for an amplification stage. At the Cs$^{137}$ peak, energy resolutions up to $\sim 3 \%$ were obtained, using a readout with Hamamatsu SiPM arrays. These results compare well with best available results obtained with a PMT readout.
Explore related subjects
Keep this discovery
M. Bonesini, R. Bertoni, R. Benocci, M. Clemenza, R. Mazza, A. deBari, A. Menegolli, M. Prata, M. Rossella. 2022-02-17. Detection of low-energy X-rays with 1/2 and 1 inch LaBr3:Ce crystals read by SIPM arrays. https://arxiv.org/abs/2202.08746
Cite the original work for its findings. Save a collection to share your selection of sources.