Filtration and breathability of nonwoven fabrics used in washable masks

Abstract

This study explores nonwoven and woven fabrics to improve upon the performance of the widespread all-cotton mask, and examines the effect of layering, machine washing and drying on their filtration and breathability for submicron and supermicron particles. Individual materials were evaluated for their quality factor, Q, which combines filtration efficiency and breathability. Filtration was tested against particles 0.5 to 5 micron aerodynamic diameter. Nonwoven polyester and nonwoven polypropylene (craft fabrics, medical masks, and medical wraps) showed higher quality factors than woven materials (flannel cotton, Kona cotton, sateen cotton). Materials with meltblown nonwoven polypropylene filtered best, especially against submicron particles. Subsequently, we combined high performing fabrics into multi-layer sets, evaluating the sets quality factors before and after our washing protocol, which included machine washing, machine drying, and isopropanol soak. Sets incorporating meltblown nonwoven polypropylene designed for filtration (Filti and surgical mask) degraded significantly post-wash in the submicron range where they excelled prior to washing (Q = 57 and 79 at 1 micron, respectively, degraded to Q = 10 and 15 post-wash). Washing caused lesser quality degradation in sets incorporating spunbond non-woven polypropylene or medical wraps (Q = 12 to 24 pre-wash, Q = 8 to 10 post-wash). Post-wash quality factors are similar for all multi-layer sets in this study, and higher than Kona quilting cotton (Q = 6). Washed multi-layer sets filtered 12-42 percent of 0.5 micron, 27-76 percent of 1 micron, 58-96 percent of 2.8 micron, and 72-100 percent of 4.2 micron particles. The measured filtration and pressure drop of both the homogeneous and heterogeneous multi-layer fabric combinations agreed with the estimations from the layering model.

Explore related subjects

Keep this discovery

BibTeXRIS

Thomas W. Bement, Ania Mitros, Rebecca Lau, Timothy A. Sipkens, Jocelyn Songer, Heidi Alexander, Devon Ostrom, Hamed Nikookar, Steven. N. Rogak. 2022-02-07. Filtration and breathability of nonwoven fabrics used in washable masks. https://arxiv.org/abs/2202.03505

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph