Gain stabilization and consistency correction approach for multiple SiPM-based gamma-ray detectors on GECAM
Abstract
Each satellite of the Gravitational wave high-energy Electromagnetic Counterpart All-sky Monitor (GECAM, mission) consists of 25 SiPM based gamma-ray detectors (GRDs). Although SiPM based GRD has merits of compact size and low bias-voltage, the drift of the SiPM gain with temperature is a severe problem for GRD performance. An adaptive voltage supply source was designed to automatically adjust the SiPM bias voltage to compensate the temperature effects and keep the gain stable. This approach has been proved to be effective during both the on-ground and in-flight tests. The in-flight measured variation of the SiPM gain is within 2%. To reduce the gain non-uniformity of GRDs, an iterative bias voltage adjustment approach is proposed and implemented. The gain non-uniformity is reduced from 17% to 0.6%. In this paper, the gain stabilization and consistency correction approach are presented and discussed in detail.
Explore related subjects
Keep this discovery
Dali Zhang, Xinqiao Li, Xiangyang Wen, Shaolin Xiong, Zhenghua An, Xilei Sun, Rui Qiao, Zhengwei Li, Ke Gong, Dongjie Hou, Yanguo Li, Xiaohua Liang, Xiaojing Liu, Yaqing Liu, Wenxi Peng, Sheng Yang, Fan Zhang, Xiaoyun Zhao, Ce Cai, Chaoyang Li, Jiacong Liu, Shuo Xiao, Chenwei Wang, Qibin Yi, Chao Zheng. 2021-10-05. Gain stabilization and consistency correction approach for multiple SiPM-based gamma-ray detectors on GECAM. https://doi.org/10.1016/j.nima.2021.166222