arXiv · 2103.05397
Extracellular electrical stimulation by ferroelectric displacement current in the switching regime
Abstract
We analyze the extracellular stimulation current and the charge injection capacitity (CIC) of microelectrodes coated with an insulating layer to prevent toxic electrochemical effects in bioelectronic applications. We show for a microelectrode coated with an insulating ferroelectric layer, that the ferroelectric polarization current contributes to the extracellular stimulation current. Depending on the remanent polarization $P_{\rm r}$ of the ferroelectric, the polarization current in the switching regime can increase the CIC by up to two orders of magnitude as compared to the commonly used extracellular capacitive stimulation with microelectrodes that are coated with a dielectric layer.
Explore related subjects
Keep this discovery
M. Becker. 2021-03-09. Extracellular electrical stimulation by ferroelectric displacement current in the switching regime. https://arxiv.org/abs/2103.05397
Cite the original work for its findings. Save a collection to share your selection of sources.