3D Printing-Assisted Energy Loss Testing of Artificial Aortic Heart Valves

Abstract

With the development of additive manufacturing, 3D printing technology has become more common nowadays. We designed and 3D printed a cost-effective liquid fluid pressure measurement device that can convert liquid pressure to air pressure. Air pressure was measured by a low-cost manometer in this study. The device can measure the pressure drop across artificial aortic heart valves at different flow rate conditions. The main objective of this study is to utilize 3D printing technology to measure pressure loss through artificial aortic heart valves. From the measured pressure drop, the energy loss of flow that occurs through the artificial aortic heart valves can be estimated. In the experiment, both mechanical and biological (bovine skin) aortic heart valves have been tested at several different flow rate conditions. By doing so, we were able to measure energy loss at different flow rates through the valves to test which type and size of the heart valve have the lowest energy loss during operation. Lower energy loss of the blood through the valve should have a lower burden to a patients heart. This study provides an example of a new application of additive manufacturing in liquid fluid pressure measurement devices and showed an acceptable performance in the experiment.

Explore related subjects

Keep this discovery

BibTeXRIS

Yuqi Jin, Akansha Rao, William Brinkman, Tae-Youl Choi. 2019-10-22. 3D Printing-Assisted Energy Loss Testing of Artificial Aortic Heart Valves. https://arxiv.org/abs/1910.11191

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph