arXiv · 1909.13706
Dielectric antenna effects in integrating line piezoelectric sensors for optoacoustic imaging
Abstract
This work studies the adverse effects, as regards noise, of immersing in water an integrating line piezoelectric detector devoted to optoacoustic imaging. We found that the sensor, in conjunction with the acoustic coupling medium (water), behaves as a resonant dielectric antenna. This phenomenon limits the performance of the system because it efficiently captures unwanted electromagnetic signals. The requirement of good acoustic coupling between the water and the sensor precluded the use of a standard metallic shielding enclosure. Therefore, we resorted to a silver-paint based electrical shield deposited on the detector. This easy-to-implement and low-cost solution significantly increases the signal to noise ratio and does not degrade the acoustic performance. The noise reduction allows the use of a better transimpedance amplifier with higher gain and bandwidth; thus achieving a very sensitive, low-noise detection system.
Explore related subjects
Keep this discovery
Roberto M. Insabella, Martin G. Gonzalez, Eduador O. Acosta, Guillermo D. Santiago. 2019-09-30. Dielectric antenna effects in integrating line piezoelectric sensors for optoacoustic imaging. https://doi.org/10.1088/1361-6501/ab9cdd
Cite the original work for its findings. Save a collection to share your selection of sources.