MRI-only brain radiotherapy: assessing the dosimetric accuracy of synthetic CT images generated using a deep learning approach

Abstract

Purpose: This study assessed the dosimetric accuracy of synthetic CT images generated from magnetic resonance imaging (MRI) data for focal brain radiation therapy, using a deep learning approach. Material and Methods: We conducted a study in 77 patients with brain tumors who had undergone both MRI and computed tomography (CT) imaging as part of their simulation for external beam treatment planning. We designed a generative adversarial network (GAN) to generate synthetic CT images from MRI images. We used Mutual Information (MI) as the loss function in the generator to overcome the misalignment between MRI and CT images (unregistered data). The model was trained using all MRI slices with corresponding CT slices from each training subject s MRI/CT pair. Results: The proposed GAN method produced an average mean absolute error (MAE) of 47.2 +- 11.0 HU over 5-fold cross validation. The overall mean Dice similarity coefficient between CT and synthetic CT images was 80% +- 6% in bone for all test data. Though training a GAN model may take several hours, the model only needs to be trained once. Generating a complete synthetic CT volume for each new patient MRI volume using a trained GAN model took only one second. Conclusions: The GAN model we developed produced highly accurate synthetic CT images from conventional, single-sequence MRI images in seconds. Our proposed method has strong potential to perform well in a clinical workflow for MRI-only brain treatment planning.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Samaneh Kazemifar, Sarah McGuire, Robert Timmerman, Zabi Wardak, Dan Nguyen, Yang Park, Steve Jiang, Amir Owrangi. 2019-04-11. MRI-only brain radiotherapy: assessing the dosimetric accuracy of synthetic CT images generated using a deep learning approach. https://doi.org/10.1016/j.radonc.2019.03.026

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Closing the gap: A two-site MLC matching study for Varian TrueBeam Linear Accelerators

Purpose: To characterise and harmonise multileaf collimator (MLC) gap calibrations across four beam-matched TrueBeams and assess a common planning-system model. Methods: Four Millennium 120 TrueBeams were studied: TS1 (installed 2016), TS5 (2021), TS6 and TS7 (2024). Gap calibration was assessed by RayStation sweeping-gap measurement, feeler gauge, the EPID-based Stakitt Fence test, and a leaf-resolved Machine Performance Check (MPC). TS1 and TS5 were adjusted in service mode, TS5 also for MLC centreline offset. Ten clinical VMAT plans, and six acquired earlier on an unadjusted reference machine, were measured with an ArcCHECK 1220 against doses computed at 0.053 and 0.065 cm x-offsets in RayStation. Results: The 2024 machines had wider baseline gaps. All four methods registered the TS1 and TS5 changes in the same direction and within 0.060 mm of predictions. After adjustment TS1 and TS5 lay within the TS6/TS7 sweeping-gap and Stakitt ranges but read wider on the feeler gauge. A conventional picket-fence test on TS5 registered no change. Raising the RayStation x-offset parameter increased mean local 3%/2 mm pass rates from 92.3% to 94.8% on TS1 and 93.2% to 95.6% on TS5; all 24 plan-criterion pairs improved on the unadjusted machine. Conclusions: Mechanical, EPID, and sweeping gap MLC gap measurements responded consistently to the MLC adjustment, moving TS1 and TS5 towards the local reference range. Every global 3%/2 mm ArcCHECK measurement exceeded the TG-218 95% limit.

physics.med-ph

Maximum-Likelihood--Based Position Decoding of Laser Processed Converging Pixel CsI: Tl Detectors for High-Resolution SPECT

This study demonstrates the feasibility of a novel fabrication technique for high spatial resolution CsI: Tl scintillation detectors tailored for single photon emission computed tomography (SPECT) systems. Building upon our previously developed laser induced optical barrier (LIOB) method, which achieved high spatial resolution, excellent sensitivity, and 100% fabrication yield in CsI: Tl detectors, we extend this approach to a converging-pixel architecture. A CsI: Tl crystal array with converging pixels was designed and fabricated, featuring entrance-face pixels of 1.6x1.6 mm2 and photodetector side pixels of 2x2 mm2. To localize gamma-ray interactions, both the center of gravity (CoG) algorithm and a maximum-likelihood (ML) based decoding method were implemented. A custom built four axis motion platform was developed to deliver a finely collimated pencil beam at precisely controlled positions and angles across the array, enabling generation of a comprehensive dataset for prior knowledge and validation. The results demonstrate an energy resolution of 11.79+/-0.53% (collimated experiment) and a position localization accuracy of 1.00+/-0.42 mm (nearest neighbor interpolation), confirming that the proposed converging-pixel architecture, combined with statistical decoding algorithms, provides a promising path toward the development of high-performance SPECT detectors.

physics.med-ph

Geant4 Simulation of an X-ray Beam Monitoring Detector based on plastic scintillators

Beam monitoring detectors play an important role in experimental control and radiological protection by enabling the real-time characterization of particle and photon beams. They provide information on beam intensity, position, and scattering, helping to ensure measurement accuracy, protect sensitive detectors, and quantify stray radiation in the surrounding environment. In this work, a Geant4 simulation was developed to investigate the performance of an X-ray beam monitoring detector (BMD) composed of six hexagonal BC404 plastic scintillators arranged to provide high granularity, using a reflector with a realistic reflectivity of 93\%. Under the simulation conditions considered in this study, the detector exhibited an intrinsic time resolution of 100.8 $\pm$ 2.0~ps. The results indicated that only about 10\% of the incident photons interacted with the detector, suggesting that the proposed design could be used for beam characterization while minimally perturbing the primary beam. In addition, the X-ray multiplicity decreased with distance according to an inverse-square dependence. A backscatter study performed using a water phantom showed that approximately 6\% of the incident beam was detected as scattered radiation. These results suggest that the proposed detector may be a useful tool for X-ray beam characterization and for studies related to scatter radiation and radiological protection. Further experimental validation is required to confirm these findings.

physics.med-ph