Abstract
The design, construction, installation, and testing of a Faraday Cup intended to measure the current of a 3 MeV, 1 microampere electron beam is described. Built as a current monitor for a M{\o}ller scattering measurement at the MIT High Voltage Research Laboratory, the device combines a large angular acceptance with the capability to measure a continuous, low energy beam. Bench studies of its performance demonstrate current measurements accurate to the percent level at 1 microampere. The Faraday Cup was designed and constructed at MIT and has been in use at the HVRL since 2017, providing a significantly more detailed measurement of beam current than was previously available.
Explore related subjects
Keep this discovery
R. Johnston, J. Bernauer, C. M. Cooke, R. Corliss, C. S. Epstein, P. Fisher, I. Friščić, D. Hasell, E. Ihloff, J. Kelsey, S. Lee, R. G. Milner, P. Moran, S. G. Steadman, C. Vidal. 2018-11-27. Realization of a Large-Acceptance Faraday Cup for 3 MeV Electrons. https://doi.org/10.1016/j.nima.2018.12.080
Cite the original work for its findings. Save a collection to share your selection of sources.