arXiv · 1711.09049
Calibration and temperature measurement of levitated optomechanical sensors
Abstract
Optically levitated nanoparticles offer enormous potential for precision sensing. However, as for any other metrology device, the absolute measurement performance of a levitated-particle sensor is limited by the accuracy of the calibration relating the measured signal to an absolute displacement of the particle. Here, we suggest and demonstrate calibration protocols for levitated-nanoparticle sensors. Our calibration procedures include the treatment of anharmonicities in the trapping potential, as well as a protocol using a harmonic driving force, which is applicable if the sensor is coupled to a heat bath of unknown temperature. Finally, using the calibration, we determine the center-of-mass temperature of an optically levitated particle in thermal equilibrium from its motion, and discuss the optimal measurement time required to determine said temperature.
Explore related subjects
Keep this discovery
Erik Hebestreit, Martin Frimmer, René Reimann, Christoph Dellago, Francesco Ricci, Lukas Novotny. 2018-04-04. Calibration and temperature measurement of levitated optomechanical sensors. https://doi.org/10.1063/1.5017119
Cite the original work for its findings. Save a collection to share your selection of sources.