arXiv · 1708.06317
Camera-limits for wide-field magnetic resonance imaging of a nitrogen-vacancy spin sensor
Abstract
Sensitive, real-time optical magnetometry with nitrogen-vacancy centers in diamond relies on accurate imaging of small ($\ll 10^{-2}$) fractional fluorescence changes across the diamond sample. We discuss the limitations on magnetic-field sensitivity resulting from the limited number of photoelectrons that a camera can record in a given time. Several types of camera sensors are analyzed and the smallest measurable magnetic-field change is estimated for each type. We show that most common sensors are of a limited use in such applications, while certain highly specific cameras allow to achieve nanotesla-level sensitivity in $1$~s of a combined exposure. Finally, we demonstrate the results obtained with a lock-in camera that pave the way for real-time, wide-field magnetometry at the nanotesla level and with micrometer resolution.
Explore related subjects
Keep this discovery
Adam M. Wojciechowski, Mürsel Karadas, Alexander Huck, Fedor Jelezko, Jan Meijer, Ulrik L. Andersen. 2017-08-21. Camera-limits for wide-field magnetic resonance imaging of a nitrogen-vacancy spin sensor. https://doi.org/10.1063/1.5010282
Cite the original work for its findings. Save a collection to share your selection of sources.