Measurement of the response of a liquid scintillation detector to monoenergetic electrons and neutrons
Abstract
The response of the liquid scintillator (EJ-301 equivalent to NE-213) to the monoenergetic electrons produced in Compton scattered $\gamma$-ray tagging has been carried out for various radioactive $\gamma$-ray sources. The measured electron response is found to be linear up to $\sim$4~MeVee and the resolution of the liquid scintillator at 1~MeVee is observed to be $\sim$~11\%. The pulse shape discrimination and pulse height response of the liquid scintillator for neutrons has been measured using $^7$Li(p,n$_1$)$^7$Be*(0.429 MeV) reaction. Non linear response to mono-energetic neutrons for the liquid scintillator is observed at E$_n$=5.3, 9.0 and 12.7 MeV. The measured response of the liquid scintillator for electrons and neutrons have been compared with Geant4 simulation.
Explore related subjects
Keep this discovery
P. C. Rout, A. Gandhi, T. Basak, R. G. Thomas, C. Ghosh, A. Mitra, G. Mishra, S. P. Behera, R. Kujur, E. T. Mirgule, B. K. Nayak, A. Saxena, Suresh Kumar, V. M. Datar. 2017-05-16. Measurement of the response of a liquid scintillation detector to monoenergetic electrons and neutrons. https://doi.org/10.1088/1748-0221%2F13%2F01%2Fp01027