Non-linearity in monochromatic transmission tomography
Abstract
While it is well known that X-ray tomography using a polychromatic source is non-linear, as the linear attenuation coefficient depends on the wavelength of the X-rays, tomography using near monochromatic sources are usually assumed to be a linear inverse problem. When sources and detectors are not treated as points the measurements are the integrals of the exponentials of line integrals and hence non-linear. In this paper we show that this non-linearity can be observed in realistic situations using both experimental measurements in a gamma-ray tomography system and simulations. We exhibit the Jacobian matrix of the non-linear forward problem. We also demonstrate a reconstruction algorithm, which we apply to experimental data and we show that improved reconstructions can be obtained over the linear approximation.
Explore related subjects
Keep this discovery
William R. B. Lionheart, Bjørn Tore Hjertaker, Rachid Maad, Ilker Meric, Sophia B. Coban, Geir Anton Johansen. 2017-05-15. Non-linearity in monochromatic transmission tomography. https://arxiv.org/abs/1705.05160