Near optimal neural network estimator for spectral x-ray photon counting data with pileup

Abstract

-Purpose: A neural network estimator to process x-ray spectral measurements from photon counting detectors with pileup. The estimator is used with an expansion of the attenuation coefficient as a linear combination of functions of energy multiplied by coefficients that depend on the material composition at points within the object [R.E. Alvarez and A. Macovski, Phys. Med. Biol., 1976, 733-744]. The estimator computes the line integrals of the coefficients from measurements with different spectra. Neural network estimators are trained with measurements of a calibration phantom with the clinical x-ray system. One estimator uses low noise training data and another network is trained with data computed by adding random noise to the low noise data. The performance of the estimators is compared to each other and to the Cramer-Rao lower bound (CRLB). Methods: The estimator performance is measured using a Monte Carlo simulation with an idealized model of a photon counting detector that includes only pileup and quantum noise. Transmitted x-ray spectra are computed for a calibration phantom. The transmitted spectra are used to compute random data for photon counting detectors with pileup. Detectors with small and large dead times are considered. Neural network training data with extremely low noise are computed by averaging the random detected data with pileup for a large numbers of exposures of the phantom. Each exposure is equivalent to a projection image or one projection of a computed tomography scan. Training data with high noise are computed by using data from one exposure. Finally, training data are computed by adding random data to the low noise data. The added random data are multivariate normal with zero mean and covariance equal to the sample covariance of data for an object with properly chosen attenuation. To test the estimators, random data are computed for different thicknesses of three test objects with different compositions. These are used as inputs to the neural network estimators. The mean squared errors (MSE), variance and square of the bias of the neural networks' outputs with the random object data are each compared to the CRLB. Results: The MSE for a network trained with low noise data and added noise is close to the CRLB for both the low and high pileup cases. Networks trained with very low noise data have low bias but large variance for both pileup cases. ralvarez@aprendtech.com Networks trained with high noise data have both large bias and large variance. Conclusion: With a properly chosen level of added training data noise, a neural network estimator for photon counting data with pileup can have variance close to the CRLB with negligible bias.

Explore related subjects

Keep this discovery

BibTeXRIS

Robert Alvarez. 2017-02-03. Near optimal neural network estimator for spectral x-ray photon counting data with pileup. https://arxiv.org/abs/1702.01006

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph