The aCORN Backscatter-Suppressed Beta Spectrometer
Abstract
Backscatter of electrons from a beta spectrometer, with incomplete energy deposition, can lead to undesirable effects in many types of experiments. We present and discuss the design and operation of a backscatter-suppressed beta spectrometer that was developed as part of a program to measure the electron-antineutrino correlation coefficient in neutron beta decay (aCORN). An array of backscatter veto detectors surrounds a plastic scintillator beta energy detector. The spectrometer contains an axial magnetic field gradient, so electrons are efficiently admitted but have a low probability for escaping back through the entrance after backscattering. The design, construction, calibration, and performance of the spectrometer are discussed.
Explore related subjects
Keep this discovery
M. T. Hassan, F. Bateman, B. Collett, G. Darius, C. DeAngelis, M. S. Dewey, G. L. Jones, A. Komives, A. Laptev, M. P. Mendenhall, J. S. Nico, G. Noid, E. J. Stephenson, I. Stern, C. Trull, F. E. Wietfeldt. 2017-01-17. The aCORN Backscatter-Suppressed Beta Spectrometer. https://doi.org/10.1016/j.nima.2017.05.029