Amplitude dependence of image quality in atomically-resolved bimodal atomic microscopy
Abstract
In bimodal FM-AFM, two flexural modes are excited simultaneously. The total vertical oscillation deflection range of the tip is the sum of the peak-to-peak amplitudes of both flexural modes (sum amplitude). We show atomically resolved images of KBr(100) in ambient conditions in bimodal AFM that display a strong correlation between image quality and sum amplitude. When the sum amplitude becomes larger than about 200 pm, the signal-to-noise ratio (SNR) is drastically decreased. We propose this is caused by the temporary presence of one or more water layers in the tip-sample gap. These water layers screen the short range interaction and must be displaced with each oscillation cycle. Further decreasing the sum amplitude, however, causes a decrease in SNR. Therefore, the highest SNR in ambient conditions is achieved when the sum amplitude is slightly less than the thickness of the primary hydration layer.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Hiroaki Ooe, Dominik Kirpal, Daniel S. Wastl, Alfred J. Weymouth, Toyoko Arai, Franz J. Giessibl. 2016-05-21. Amplitude dependence of image quality in atomically-resolved bimodal atomic microscopy. https://doi.org/10.1063/1.4964125