Study of the dependency on magnetic field and bias voltage of an AC-biased TES microcalorimeter
Abstract
At SRON we are studying the performance of a Goddard Space Flight Center single pixel TES microcalorimeter operated in an AC bias configuration. For x-ray photons at 6 keV the pixel shows an x-ray energy resolution $dE_{FWHM} = 3.7\mathrm{eV}$, which is about a factor 2 worse than the energy resolution observed in an identical DC-biased pixel. In order to better understand the reasons for this discrepancy we characterised the detector as a function of temperature, bias working point and applied perpendicular magnetic field. A strong periodic dependency of the detector noise on the TES AC bias voltage is measured. We discuss the results in the framework of the recently observed weak-link behaviour of a TES microcalorimeter
Explore related subjects
Keep this discovery
Luciano Gottardi, James Adams, Catherine Bailey, Simon Bandler, Marcel Bruijn, James Chervenak, Megan Eckart, Frank Finkbeiner, Roland den Hartog, Henk Hoevers, Richard Kelley, Caroline Kilbourne, Piet de Korte, Jan van der Kuur, Mark Lindeman, Frederick Porter, Jack Sadleir, Steve Smith. 2016-04-09. Study of the dependency on magnetic field and bias voltage of an AC-biased TES microcalorimeter. https://doi.org/10.1007/s10909-012-0494-x