arXiv · 1509.09021
A 128-channel picoammeter system and its application on charged particle beam current distribution measurements
Abstract
A 128-channel picoammeter system is constructed based on instrumentation amplifiers. Taking the advantages of high electric potential and narrow bandwidth in DC energetic charged beam measurements, current resolution better than 5 fA can be achieved. Two 128-channel Faraday cup arrays are built, and are employed for ion and electron beam current distribution measurements. Tests with 60 keV O3+ ions and 2 keV electrons show that it can provide exact boundaries when a positive charged particle beam current distribution is measured.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Deyang Yu, Junliang Liu, Yingli Xue, Mingwu Zhang, Xiaohong Cai, Jianjun Hu, Jinmei Dong, Xin Li. 2015-09-30. A 128-channel picoammeter system and its application on charged particle beam current distribution measurements. https://doi.org/10.1063/1.4934849