High-resolution ion pulse ionization chamber with air filling for the Rn-222 decays detection
Abstract
The construction and characteristics of the cylindrical ion pulse ionization chamber (CIPIC) with a working volume of 3.2 L are described. The chamber is intended to register alpha-particles from the $^{222}$Rn and its daughter's decays in the filled air sample. The detector is less sensitive to electromagnetic pick-ups and mechanical noises. The digital pulse processing method is proposed to improve the energy resolution of the ion pulse ionization chamber. An energy resolution of 1.6% has been achieved for the 5.49 MeV alpha-line. The dependence of the energy resolution on high voltage and working media pressure has been investigated and the results are presented.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Yu. M. Gavrilyuk, A. M. Gangapshev, A. M. Gezhaev, R. A. Etezov, V. V. Kazalov, V. V. Kuzminov, S. I. Panasenko, S. S. Ratkevich, D. A. Tekueva, S. P. Yakimenko. 2015-08-18. High-resolution ion pulse ionization chamber with air filling for the Rn-222 decays detection. https://doi.org/10.1016/j.nima.2015.08.034