Operation of the CESR-TA vertical beam size monitor at $E_{\rm b}$=4 GeV
Abstract
We describe operation of the CESR-TA vertical beam size monitor (xBSM) with $e^\pm$ beams with $E_{\rm b}$=4 GeV. The xBSM measures vertical beam size by imaging synchrotron radiation x-rays through an optical element onto a detector array of 32 InGaAs photodiodes with 50 $\mu$m pitch. The device has previously been successfully used to measure vertical beam sizes of 10-100 $\mu$m on a bunch-by-bunch, turn-by-turn basis at $e^\pm$ beam energies of $\sim$2 GeV and source magnetic fields below 2.8 kG, for which the detector required calibration for incident x-rays of 1-5 keV. At $E_{\rm b}=4.0$ GeV and $B$=4.5 kG, however, the incident synchrotron radiation spectrum extends to $\sim$20 keV, requiring calibration of detector response in that regime. Such a calibration is described and then used to analyze data taken with several different thicknesses of filters in front of the detector. We obtain a relative precision of better than 4% on beam size measurement from 15-100 $\mu$m over several different ranges of x-ray energy, including both 1-12 keV and 6-17 keV. The response of an identical detector, but tilted vertically by 60$^\circ$ in order to increase magnfication without a longer beamline, is measured and shown to improve x-ray detection above 4 keV without compromising sensitivity to beam size. We also investigate operation of a coded aperture using gold masking backed by synthetic diamond.
Explore related subjects
Keep this discovery
J. P. Alexander, C. Conolly, E. Edwards, J. W. Flanagan, E. Fontes, B. K. Heltsley, A. Lyndaker, D. P. Peterson, N. T. Rider, D. L. Rubin, R. Seeley, J. Shanks. 2015-07-09. Operation of the CESR-TA vertical beam size monitor at $E_{\rm b}$=4 GeV. https://doi.org/10.1016/j.nima.2015.07.028
Cite the original work for its findings. Save a collection to share your selection of sources.