arXiv · 1311.3265
Current Sensing Noise Thermometry: A fast practical solution to low temperature measurement
Abstract
We describe the design and performance of a series of fast, precise current sensing noise thermometers. The thermometers have been fabricated with a range of resistances from 1.290 $\Omega$ down to 0.2 m$\mathrm{\Omega}$. This results in either a thermometer that has been optimised for speed, taking advantage of the improvements in superconducting quantum interference device (SQUID) noise and bandwidth, or a thermometer optimised for ultra-low temperature measurement, minimising the system noise temperature. By using a single temperature calibration point, we show that noise thermometers can be used for accurate measurements over a wide range of temperatures below 4 K. Comparisons with a melting curve thermometer, a calibrated germanium thermometer and a pulsed platinum nuclear magnetic resonance thermometer are presented. For the 1.290 $\mathrm{\Omega}$ resistance we measure a 1 % precision in just 100 ms, and have shown this to be independent of temperature.
Explore related subjects
Keep this discovery
Andrew Casey, Frank Arnold, Lev V. Levitin, Chris P. Lusher, John Saunders, Aya Shibahara, Harriet van der Vliet, Dietmar Drung, Thomas Schurig, Graham Batey, Michael Cuthbert, Anthony Matthews. 2013-11-13. Current Sensing Noise Thermometry: A fast practical solution to low temperature measurement. https://doi.org/10.1007/s10909-014-1147-z
Cite the original work for its findings. Save a collection to share your selection of sources.