arXiv · 1307.7373
Single-reference high-precision mass measurement with a multi-reflection time-of-flight mass spectrograph
Abstract
A multi-reflection time-of-flight mass spectrograph, competitive with Penning trap mass spectrometers, has been built at RIKEN. We have performed a first online mass measurement, using 8Li+ (T1/2 = 838 ms). A new analysis method has been realized, with which, using only 12C+ references, the mass excess of 8Li was accurately determined to be 20947.6(15)(34) keV (dm/m = 6.6 x 10-7). The speed, precision and accuracy of this first online measurement exemplifies the potential for using this new type of mass spectrograph for precision measurements of short-lived nuclei.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Y. Ito, P. Schury, M. Wada, S. Naimi, T. Sonoda, H. Mita, F. Arai, A. Takamine, K. Okada, A. Ozawa, H. Wollnik. 2013-07-28. Single-reference high-precision mass measurement with a multi-reflection time-of-flight mass spectrograph. https://doi.org/10.1103/physrevc.88.011306