arXiv · 1002.0275
High resolution magnetostriction measurements in pulsed magnetic fields using Fibre Bragg Gratings
Abstract
We report on a new high resolution apparatus for measuring magnetostriction suitable for use at cryogenic temperatures in pulsed high magnetic fields which we have developed at the Hochfeld-Magnetlabor Dresden. Optical fibre strain gauges based on Fibre Bragg Gratings are used to measure the strain in small (~1mm) samples. We describe the implementation of a fast measurement system capable of resolving strains in the order of $10^{-7}$ with a full bandwidth of 47kHz, and demonstrate its use on single crystal samples of GdSb and GdSi.
Explore related subjects
Keep this discovery
Ramzy Daou, Franziska Weickert, Michael Nicklas, Frank Steglich, Ariane Haase, Mathias Doerr. 2010-02-01. High resolution magnetostriction measurements in pulsed magnetic fields using Fibre Bragg Gratings. https://doi.org/10.1063/1.3356980
Cite the original work for its findings. Save a collection to share your selection of sources.