arXiv · 0704.2990
The principle of a virtual multi-channel lock-in amplifier and its application to magnetoelectric measurement system
Abstract
This letter presents principles and applications of a virtual multi-channel lock-in amplifier that is a simple but effective method to recover small ac signal from noise with high presison. The fundamentals of this method are based on calculation of cross-correlation function. Via this method, we successfully built up a magnetoelectric measurement system which can perform precise and versatile measurements without any analog lock-in amplifier. Using the virtual multi-channel lock-in amplifier, the output of the magnetoelectric measurement system is extensively rich in magnetoelectric coupling behaviors, including coupling strength and phase lag, under various dc bias magnetic field and ac magnetic field.
Explore related subjects
Keep this discovery
Jun Lu, De-An Pan, Lijie Qiao. 2007-04-23. The principle of a virtual multi-channel lock-in amplifier and its application to magnetoelectric measurement system. https://doi.org/10.1088/0957-0233/19/4/045702